31 VWR Mega Star 600 / 600R ver b, rel 17.07.2020
Troubleshooting by Guide
If problems occur other than those listed in this table, the authorized customer service representative must
be contacted.
Error number Error message Troubleshooting
E-002;
E-005;
E-008;
E-010;
E-011;
E-012;
E-015;
E-016;
E-034;
E-036;
E-041;
E-048;
E-050;
E-051;
E-052;
E-053;
E-054;
E-072;
E-077;
E-101;
E-104
Read Manual Restart the centrifuge.
If the error message appears again, inform the service.
E-031 Temp High! CAUTION
Hot metal parts!
Check, if the centrifuge is accessible.
Be sure, that the room temperature is within the limits.
Let the centrifuge cool down for 15 minutes.
Be sure there is no condensed water in the rotor chamber.
If the error message appears again, inform the service.
E-017;
E-020;
E-021;
E-022;
E-023
E-078;
E-079;
E-080;
E-081;
Read Manual Wait until the rotor has stopped.
Check, if the rotor is qualified for the VWR Mega Star 600 / 600R
Centrifuge.
Check, if the bottom of the rotor is damaged and if the rotor is placed
on the Auto-Lock correctly.
If the error message appears again, inform the service.