Stylus does not register
icons on the display
Touch screen
display requires
calibration
Select the calibration
check icon and choose
touch screen, and follow
the instructions.
Icons missing from the
display
Icon has been
de-selected
Hold stylus on to the
blue area of the display,
a list will appear, ensure
required icon is selected.
Calibration procedure
failed or cannot be
completed
Turbine may be
faulty
Repeat calibration
procedure, if problem
still persists, replace
turbine or return unit for
servicing
Turbine not
tted tightly
to calibration
syringe
Ensure the syringe is
tted to the turbine using
the adapter supplied part
# 861427 (supplied with
the syringe)
Calibration
syringe does not
have an
inspiratory seal
or seal is leaking.
Ensure you are using a
Vyaire syringe.
Shaft of the
syringe being
pushed down
The syringe should be
emptied and lled with
one smooth stroke, avoid
pushing down on the
shaft or banging at the
end of each manoeuver.
39