EasyManua.ls Logo

Waters ACQUITY CH-A - Page 23

Waters ACQUITY CH-A
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Required tools and materials
Chemical-resistant, powder-free gloves
Protective eyewear
1/4-inch open-end wrench
Needle-nose pliers
Flat-blade screwdriver
T20 TORX driver
To prepare the CH-A for installation with the MS-bracket kit:
1. Stop any running mobile phase.
2. Power-off the sample manager.
3. If a column and APH or column stabilizer assembly are installed within the column
compartment, complete steps 4 through 9 to remove them from the compartment.
4. Detach the column's eCord fob from the receptacle on the right-hand side of the column
heater.
5. Using the 1/4-inch open-end wrench, disconnect the long, flat compression-screw end of
the APH assembly from port 6 of the sample manager's injection valve.
Figure 2–4: Removing APH assembly from the sample manager injector valve
1
Long, flat compression-screw end of the APH assembly
6. Disconnect the column-outlet PEEK tubing from the detector inlet.
7. After allowing sufficient time for cooling, open the column compartment door, and then
disconnect and remove the column (and the column in-line filter, if fitted) from the APH
assembly.
Note: Loosen the #2 locking cap nut on the APH assembly before loosening the #1
compression fitting.
December 4, 2019, 715005052 Version 02 (previously released as Rev.B)
Page 23

Table of Contents

Related product manuals