EasyManua.ls Logo

Waters ACQUITY CH-A - Page 71

Waters ACQUITY CH-A
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
To replace the leak sensor:
1. Power-off the sample manager.
2. Depending on the configuration of the column heater leak-sensor you are replacing, do as
follows:
If the leak sensor is in the sample manager's upper drip tray, follow the instructions for
replacing the leak sensor in the ACQUITY UPLC Sample Manager - Flow Through
Needle PLUS Series Overview and Maintenance Guide.
If the leak sensor is in the MS-bracket drip tray, proceed with the steps below, to
replace the leak sensor.
3. Disconnect the APH-inlet tubing from injector port 6 of the sample manager.
4. Disconnect the column-outlet tubing from the detector inlet.
5. Disconnect the eCord fob from its receptacle on the right-hand side of the CH-A.
6. Using the T20 TORX driver, remove the TORX screw from the right-hand side of the
column heater, behind the eCord receptacle.
Figure 3–1: TORX screw behind the eCord receptacle
TORX screw
eCord receptacle
7. Gently pull the right-hand side of the column heater toward you, and then slide it to the
right, to unseat the tab on the left-hand side of the column heater from its extended
column-heater housing.
8. Using the flat-blade screwdriver, loosen the two retaining screws, and then remove the CH-
A interconnect cable's 25-pin connector from the D-sub receptacle on the rear of the
column heater.
December 4, 2019, 715005052 Version 02 (previously released as Rev.B)
Page 71

Table of Contents

Related product manuals