EasyManua.ls Logo

Welch Allyn MicroTymp 2 - Troubleshooting the MicroTymp 2 Handle; Functional Checks of the Printer;Charger

Welch Allyn MicroTymp 2
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-53
Appendix D
2
4
3
2
Troubleshooting the MicroTymp2 Handle
Symptom Possible Cause Possible Solution
Handle Does No battery. Put battery in.
Not Turn On
Battery in backwards. Reposition battery
observing polarity.
Battery not charged/dead. Charge/replace battery.
If symptom still persists,
return to local Welch Allyn
service location for service.
Too Much Too much movement See page A-31.
Artifact on LCD during test.
Handle has too much Check handle in cavity
internal noise. (page D-51). If handle
passes cavity test,
artifact is due to
motion. If handle does not
pass cavity test, return
to Welch Allyn service
location for service.
“Frozen” Display Microcomputer has Push TEST and repeat.
on LCD malfunctioned.
If symptom persists, push
all three handle buttons
(TEST,R MEM, L MEM)
simultaneously to reset
OR microcomputer.
If symptom persists, remove
and reinsert the battery.
If symptom persists, return
“Checkerboard” to local Welch Allyn service
Pattern on LCD location for service.
Functional Checks of the Printer/Charger
Pressing the FEED and PRINT buttons simultaneously causes a
test pattern to print. See Figure 45 (on page D-54).
Test Pattern
The test pattern is used to confirm that the print head is functioning
properly. If any of the print head elements are not functioning, a white
line will appear vertically down the printout. A defect in the paper may
also cause white lines or light printing. Repeat the test pattern to confirm
any suspected printing problems. If paper advances but nothing prints
out, check to be sure the paper is inserted properly (see page C-48).
1

Table of Contents

Related product manuals