Restoring and restarting system messages
Message Text Meaning
Are you sure you want to
delete the custom loaded
criteria and restore system
default criteria?
To save your custom
criteria, use the Export
option in Import/Export
tool.
You have requested a restore to the factory default settings for the
age based criteria settings. If you select Restore, current criteria
settings are removed.
The database has been reset
and recovery files have been
built. (selecting export
from the tools menu will
transfer the recovery
files).
There was an internal error while restoring the subject history file.
A backup recovery file has been generated and copied to the
recovery area for export. You can select Delete to delete the
recovered files or Cancel to retain the files. If the recovered files
remain on the device, this message displays each time you restart
the device until you export the files.
Sorry for the inconvenience.
We need to reboot the
device.
Screening mode has encountered a problem and the device needs
to be restarted. You can restart the device or, if you select Cancel,
continue in Read Only mode. Read Only mode lets you review
existing screening results. You cannot perform new screenings until
you restart the device.
Problem encountered. The
device needs to reboot.
The system was not able to start and has attempted to correct the
issue. Restart the device.
Un-calibrated device The system has not been able to read the device configuration file.
Restart the device.
For more information on troubleshooting or error messages, contact Welch Allyn
Technical Support: www.welchallyn.com.
Directions for use Troubleshooting 43