Operation pH 330i/340i
16
AutoRead
(drift control)
The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.
For identical measurement conditions, the following criteria apply:
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
Reproducibility Response time
Better than 0.02 > 30 seconds
1 Call up the pH measuring mode with <M> and <> <> .
2 Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3 Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4 If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5 To terminate the AutoRead function: Press the <AR> key.