EasyManua.ls Logo

wtw 330i - Instrument Calibration

wtw 330i
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation pH 330i/340i
18
4.3 Calibrating
Why calibrate? pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is dis-
played. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the measuring instru-
ment.
Thus, you should calibrate at regular intervals.
When to calibrate?
l After connecting another electrode
l When the sensor symbol flashes:
after the calibration interval has expired
after a voltage interruption, e. g. after changing the batteries
Calibration interval The calibration interval (Int 3) is set to 7 days in the factory. You can
select the interval in the range of 1 ... 999 days (see section
4.6 C
ONFIGURATION).
You can select between 3 calibration procedures:
AutoCal TEC is specially adapted to the WTW technical buffer solutions as a fully au-
tomatic two-point or three-point calibration. The buffer solutions are
automatically recognized by the measuring instrument. Depending on
the instrument setting (see section 4.6 C
ONFIGURATION), the instrument
displays the relevant buffer nominal value or the current electrode volt-
age in mV. The calibration can be terminated after the first buffer solu-
tion. This corresponds to a single-point calibration. For this, the
instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and deter-
mines the asymmetry of the electrode.
AutoCal DIN is specially adapted to the permanently programmed buffer solutions in
accordance with DIN19266 as a fully automatic two-point or three-
point calibration. The buffer solutions are automatically recognized by
the measuring instrument. Depending on the instrument setting (see
section 4.6 C
ONFIGURATION), the instrument displays the relevant buffer
nominal value or the current electrode voltage in mV. The calibration
can be terminated after the first buffer solution. This corresponds to a
single-point calibration. For this, the instrument uses the Nernst
slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the
electrode.
ConCal This function is a conventional two-point calibration using two buffer
solutions (pH 7.0 ± 0.5 and any other buffer solution) or a single-point
calibration using any buffer solution and is used as a high-speed
method.

Related product manuals