Operation pH 330i/340i
32
4.4 Saving
The measuring instrument has an internal data memory. It can store up
to 800 data records.
A complete data record consists of:
l Storage location
l Date/time
l Measured value
l Temperature
l Temperature measuring procedure
l ID number
You can transmit measured values (data records) to the data storage
in two ways:
l Save manually
l Switch on AutoStore (Int 1).
4.4.1 Saving manually
You can transmit a measured value to the data storage as follows:
1 Press the <STO> key.
The current number (location number No.) of the next free stor-
age location appears under the current measured value on the
display.
2 Confirm with <RUN/ENTER>.
The display switches to entering the ID number.
Ident
°C
2
4
TP
mV/pH
mol/L
8
Auto
TempError
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCal TEC
ConCal
Time
Baud
Day.Month
Year
1
U
%
°F
S
mg/L
% ppm
ISE
4
29
9
6
pH
Store