EasyManua.ls Logo

wtw 330i - AutoCal DIN Calibration

wtw 330i
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
pH 330i/340i Operation
25
4.3.2 AutoCal DIN
For this procedure, use two or three different standard buffer solutions
according to DIN 19266 in ascending or descending order (type A, C,
D or F with pH values at 25 °C of: 1.679 / 4.006 / 6.865 / 9.180).
Note
The buffer solutions are automatically recognized by the measuring in-
strument. Depending on the instrument setting (see section
4.6 C
ONFIGURATION), the instrument displays the relevant buffer nomi-
nal value or the current electrode voltage in mV.
Note
Skip the steps 2, 7 and 14 if you use a temperature sensor.
1 Press the <CAL> key repeatedly until the display Cd1 and the
function display AutoCal DIN appear. The sensor symbol dis-
plays the evaluation of the last calibration (or no sensor symbol
in the delivery state or after the measurement parameter has
been reset).
2 If necessary, enter the temperature of the buffer solution with
<▲> <▼>.
3 Immerse the pH electrode in the first buffer solution.
4 Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:
AutoCal TEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
1C
d
°C
TP
pH
AutoCal DIN