EasyManua.ls Logo

wtw 330i - Page 24

wtw 330i
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation pH 330i/340i
24
Three-point calibration
Note
You can prematurely terminate the three-point calibration by pressing
<M>. The values of the two-point calibration for slope and asymmetry
will then remain stored.
12 Press the <RUN/ENTER> key
to continue the three-point calibration.
Ct3 appears on the display.
13 Thoroughly rinse the electrode with distilled water.
14 If necessary, enter the temperature of the third buffer solution
with <▲> <▼>.
15 Immerse the pH electrode in the third buffer solution.
16 Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display.
When the measured value is stable, AR disappears.
The instrument displays the value of the slope (mV/pH) and the
evaluation of the calibration in calibration range II.
17 Press the <RUN/ENTER> key.
The instrument displays the value of the asymmetry in calibra-
tion range II. The three-point calibration is completed.
18 To return to the measuring mode: Press the <M> key.