pH 330i/340i Operation
43
Setting the
Data transmission
interval
The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min,
5 min, 10 min, 15 min, 30 min, 60 min):
1 Press the <RUN/ENTER> key and hold it down.
2 Press the <RCL> key. Int 2 appears on the display.
3 Set the required interval between the saving procedures with
<> <>.
4 Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pH-
measuring mode.