pH 330i/340i What to do if...
55
Obviously incorrect
measured values
Instrument does not
react to keystroke
Display to
toto
to
Probe symbol flashes
Message StoFull
StoFullStoFull
StoFull
Cause Remedy
pH electrode:
pH electrode unsuitable Use suitable electrode
Temperature difference
between buffer and test
sample too high
Adjust temperature of buffers
or test samples
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Operating condition
undefined or EMC load
unallowed
Processor reset:
Press the <CAL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Cause Remedy
Time-out of the interface Check the instrument that is
connected
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
All memory locations are full Output data storage and clear
data storage