Cond 7110 Conductivity
ba75927e03 04/2012 21
3. Confirm the selection of (Use Cell) with <ENTER> or <CAL__>.
The cell constant of the last calibration is displayed.
4. If necessary, select the suitable cell constant with <CAL>.
5. Use <ENTER> to finish the calibration.
The selected calibrated cell constant is used.
5.2.5 Setting the cell constant manually
This method can be used when the cell constant of the measuring cell is known
(e.g. taken from the operating manual of the measuring cell, or from the labeling
on the measuring cell).
1. Open the calibration menu with <CAL>.
2. If necessary, select the calibration type (Use Cell) with <CAL>.
Use Cell is displayed.
L
CL
E
E
S
U
1/cm
9
0 7
4
L
AC
Cal