Conductivity Cond 7110
22 ba75927e03 04/2012
3. Confirm the selection of (Use Cell) with <ENTER> or <CAL__>.
The cell constant that was set last is displayed.
4. Select a cell constant in a suitable range with <CAL>.
• Range 0.250 ... 2.500 cm-1
• Range 0.090 ... 0.110 cm-1
• Permanent, 0.010 cm
-1
5. Set the exact cell constant within the range with <><> as necessary.
6. Use <ENTER> to finish the calibration.
The adjusted cell constant is used.
5.2.6 Calibration data
Displaying the calibration data
1. Show the calibration data in the measured value display with <CAL__>.
The calibrated cell constant is displayed.