What to do if... Cond 7110
32 ba75927e03 04/2012
9 What to do if...
9.1 Conductivity
9.1.1 Error message E3
9.1.2 Error message OFL, UFL
The measured value is outside the measuring range.
9.2 General information
9.2.1 [LoBat] Display
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Measuring cell contaminated Clean the measuring cell and
replace it if necessary
Calibration solution not suitable Check the calibration solutions
Cause Remedy
Measured value outside the
measuring range
Use suitable conductivity
measuring cell
Cause Remedy
Batteries almost empty Replace the batteries (see section
3.3.1 I
NSERTING THE BATTERIES,
page 10)