Cond 7110 Reset
ba75927e03 04/2012 29
Measurement and system settings that can be reset
The following settings are reset to the delivery condition
(default):
Measurement settings
System settings
Calibration data
The measuring system is possibly not calibrated after a reset.
Before measuring, make sure the meter uses the cell con-
stant suitable for the measuring cell.
Measurement settings Default
Reference temperature (Tref)t25
Unit of the measured temperature
value (Uni )
°C
Calibration interval (Int.C ) 180 d
Temperature compensation nLF
Temperature coefficient of the linear
temperature compensation
2.000 %/K
TDS factor 1.00
Manually set cell constant
0.475 cm
-1
System settings Default
Switch-off interval (.Off )1 h