What to do if... pH 3110
40
ba75791e02 11/2008
No stable measured
value
Sensor symbol flashes
Display,
LoBat
Cause Remedy
pH combination electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if
necessary
Combination electrode + test
sample:
Conductivity too low Use suitable combination
electrode
Temperature too high Use suitable combination
electrode
Organic liquids Use suitable combination
electrode
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Batteries almost empty Replace the batteries (see
section 5.1 M
AINTENANCE)