wtw pH 3110 - Incorrect Measured Values; Unresponsive Keystrokes; Software Version Check

wtw pH 3110
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
pH 3110 What to do if...
41
ba75791e02 11/2008
Obviously incorrect
measured values
Meter does not react to
keystroke
You want to know which
software
version is in the meter
Cause Remedy
pH combination electrode:
– pH combination electrode
unsuitable
– Use suitable combination
electrode
– Temperature difference
between buffer and test sample
too high
– Adjust temperature of buffer
or sample solutions
– Measurement procedure not
suitable
– Follow special procedure
Cause Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <ENTER> and
<On/Off>
key
simultaneously
Cause Remedy
– E. g., a question by the service
department
– Switch on the meter.
During the display test,
display the software version
with <ENTER>.