EasyManuals Logo

wtw Turb 2000 User Manual

wtw Turb 2000
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
Turb 2000 Series Installation and commissioning
ba75695e03 10/2013 11
2.2 Plumbing
Figure 4 shows the plumbing recommended for the instrument. The instrument requires only a very low
minimum pressure for operation. A flow controller and backpressure valve are integrated in the
instrument. For details, see chapter Technical data.
IN
2
33
8
9
10
6
7
1
5
4
12
14
11
13
Figure 4: Recommended plumbing for the instrument
1 Feed tubing connection
2 Shutoff clamp
3 Feed tubing connection
4 Backpressure valve
5 Feed tubing connection
6 Drain tubing connection
7 Backpressure valve
8 Drain tubing connection
9 Drain vent
10 Drain tubing connection
11 Sensor connection cable
12 Insert
13 Tubing (not supplied)
14 Nut
The instrument is equipped to be plumbed using 4.75 mm (3/16”) ID, 8 mm (5/16”) OD flexible tubing.
Opaque tubing should be used if the tubing will be exposed to sunlight, to prevent algae growth. For
connectiong we recommend the conection set with quick connector (A-Set-DW, siehe section 11).
In figure 4, there are three flow devices shown. On the inlet side there is a shutoff clamp (Pos. 2) used
during cuvette maintenance, and the flow controller (Pos. 4). With the backpressure valve (Pos. 7) a
back pressure is created in the cell. Backpressure may be required to prevent air from coming out of
solution, which may be observed as tiny air bubbles.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Turb 2000 and is the answer not in the manual?

wtw Turb 2000 Specifications

General IconGeneral
Brandwtw
ModelTurb 2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals