Maintenance
9
READING LAMP REPLACEMENT (P/N 301-21)
Replace the reading area lamp as follows:
1. Remove the unit bottom cover by removing the four (4) screws located
in the rubber feet.
2. Remove the four (4) screws securing the Lamp P.C. board to the arm
assembly at bottom of chassis.
3. Disconnect lamp connector and remove faulty Lamp P.C. board.
NOTE: Ensure lamp P.C. board bulb is clean. Use alcohol on a
lint free cloth or cotton swab to remove smudges.
CAUTION: The bulb on Lamp P.C. board has been prealigned
for optimum optical performance. Use care not to
bend or reposition the bulb when cleaning or
installing.
4 Connect the lamp connector to new Lamp P.C. board.
5. Secure Lamp P.C. board to arm assembly with four (4) screws.
6. Reinstall bottom cover and secure with four (4) screws.
Figure 7. Lamp PC Board Access
Screws
Screws
Lamp PC
Board Assy
Lamp Connector