What to do if my X-Rite DigiMatch PPG show measurement error?
- VvaughannathanielAug 3, 2025
If your device shows measurement errors or inaccurate results, it could be due to several reasons. First, the material being measured might be damaged; try repeating the measurement. Second, the calibration might have been performed on a dirty white tile or camera test target; clean the calibration reference and repeat the calibration. The device itself might require calibration; perform a health check and calibration on the calibration reference. If these steps fail, the device might be defective, and you should contact technical support.