Section 1. Features and Specifications
Electrical Specifications
14 Operating Manual for HPD Series Power Supply
Electrical Specifications
Specifications are warranted over a temperature range of 0 to 30 °C with default
local sensing. Above 30 °C, the output voltage needs to be derated linearly to zero at
70 °C. Specifications are subject to change without notice.
Table 1.2 Electrical Specifications for 15 V to 60 V Models
Models 15–20 30–10 60–5
Output Ratings:
Output Voltage
Output Current
Output Power
015 V
020 A
300 W
030 V
010 A
300 W
060 V
05A
300 W
Line Regulation:
1
Voltage (0.01% of Vmax + 2 mV)
Current (0.01% of Imax + 1 mA)
3.5 mV
4mA
5mV
3mA
8mV
2.5 mA
Load Regulation:
2
Voltage (0.01% of Vmax + 2 mV)
Current (0.01% of Imax + 1 mA)
3.5 mV
4mA
5mV
3mA
8mV
2.5 mA
Meter Accuracy:
Voltage (1% of Vmax + 1 count)
Current (1% of Imax + 1 count)
250 mV
300 mA
400 mV
200 mA
700 mV
60 mA
Output Noise and Ripple
(20 Hz-20 MHz):
Voltage (p-p)
Voltage (rms)
Current (rms)
3
100 mV
5mV
20 mA
100 mV
5mV
10 mA
100 mV
5mV
5mA
Drift (60 minutes):
4
Voltage (0.02% of Vmax)
Current (0.03% of Imax)
3mV
6mA
6mV
3mA
12 mV
1.5 mA
Drift (8 hours):
5
Voltage (0.02% of Vmax)
Current (0.08% of Imax)
3mV
16 mA
6mV
8mA
12 mV
4mA
Temperature Coefficient
6
Voltage (0.015% of Vmax/°C)
Current (0.02% of Imax/°C)
2.25 mV
4mA
4.5 mV
2mA
9mV
1mA
1. For input voltage variation over the AC input voltage range, with constant rated load.
2. For 0 to 100% load variation, with constant nominal line voltage.
3. Current mode noise is measured from 10% to 100% of rated output voltage, full current.
4. Maximum drift over 60 minutes with constant line, load, and temperature, after 60-minute
warm-up.
5. Maximum drift over 8 hours with constant line, load, and temperature, after 60-minute
warm-up.
6. Change in output per °C change in ambient temperature, with constant line and load.