Release 1.2 33
Section 4. Operation
Introduction
Once you have installed the power supply and have connected both the AC input
power and the load as covered in Section 2, “Installation”, the power supply is ready
to operate.
“Operating Modes”, below, offers a brief explanation of Constant Voltage and
Constant Current Mode operation.
“Connecting Multiple Supplies in Series (Voltage Mode Only)” on page 36,
covers using multiple supplies.
Operating Modes
Your power supply has two basic operating modes: Constant Voltage Mode and
Constant Current Mode. The mode in which the power supply operates at any given
time depends on the combination of:
Output voltage setting V
SET
Output current limit setting I
SET
Resistance or impedance of the attached load R
L
Figure 4.1 represents the relationships between these variables.
Figure 4.1 Relationship Between Operating Mode Variables
Output
Current
Output
Voltage
Crossover Point
Constant Current
Mode Region
O
V SET
I SET
O
Constant Voltage
Mode Region
Where:
R
L
= Load Resistance
V
SET
= Output Voltage Setting
I
SET
= Output Current Setting
R
L
=
V
SET
I
SET
R
L
<
V
SET
I
SET
R
L
>
V
SET
I
SET
Note
The control circuits have been designed to allow you to set output voltage and current
up to 5% over the model-rated maximum values. The power supply will operate within
these extended ranges, but we cannot guarantee full performance to specification.