89
Ambient conditions:
Do not use the device in hazardous locations!
Climate: Ambient temperature: + 10 ... + 40 °C for operation and storage
Humidity according to EN 61 010, Part 1:
Max. relative humidity 80 % for temperatures up to 31 °C,
linear decrease down to 50 % relative humidity at a temperature of 40 °C
Interchangeable units:
Compatibility: units are compatible to the:
- the titrators TitroLine
®
7000 / 7500 / 7500 KF / 7750 / 7800
- the Piston Burette TITRONIC
®
500
Recognition: automatically through RFID
Recognition of unit size and characteristics of the Titration- or dosing solution
Valve: volume neutral cone valve made from fluorocarbon polymers (PTFE), TZ 3000
Cylinder: borosilicate glass 3.3 (DURAN
®
)
Hoses: FEP hose set, blue
Bracket for
supply bottle: suitable for square glass bottle and misc. reagent bottles
Materials: borosilicate glass DURAN
®
, fluorocarbon polymers (PTFE), stainless steel, polypropylene
Dimensions: 15,3 x 45 x 29,6 cm (W x H x D) incl. reagent bottle
Weight: approx. 1.2 kg for interchangeable unit WA incl. empty reagent bottle
Dosing accuracy:
after DIN EN ISO 8655, part 3:
Accuracy: 0.15 %
Precision: 0.05 - 0.07 %
(in dependence of the used interchangeable unit)
Dosing accuracy of the piston burette TITRONIC
®
500 with interchangeable units (WA):
Interchangeable
unit type No.
Tolerances of the Ø
i
of the glass cylinder [mm]
Dosage error according to
100 % volume [%]