Command Description Reply
aaAA automatic allocation of device address aaY
aaMC1...XX choosing a method aaY
aaBF „filling burette“. Aufsatz wird gefüllt. aaY
aaBV output of dosed volume in ml aa0.200
aaDA dose volume without filling, with adding the volume aaY
aaDB dose volume without filling, reset of the volume aaY
aaDO dose volume with filling, without adding the volume aaY
aaGDM dosing speed in ml/min aaY
aaGF filling time in seconds (min is 20, default 30) aaY
aaES „ESC" function one step backwards aaY
aaEX „EXIT“ function.back to main menu aaY
aaGDM dosing speed in ml/min (0.01 – 100 ml/min) aaY
aaGF filling time in sec (adjustable 20 – 999 seconds) aaY
aaGS output serial no. Of device aaGS08154711
aaLR output report (short report) aaY
aaLI output method content
aaLO output documentation (as configured)
aaRH request of identification aaIdent: TITRONIC
®
500
aaRC send last command aa“last command“
aaRS report status aaStatus:“text
possible answers are:
„STATUS:READY“ for ready
„STATUS:dosing“ dosing
„STATUS:filling“ filling
„ERROR:busy“ if no interchangeable unit has been attached
aaSM start selected method aaY
aaSEEPROM EEPROM reset to factory defaults aaY
aaSR stop the actual function aaY
aaSYS5 adjust language to „German“ aaY
aaSYS1 adjust language to „English“ aaY
aaSYS2 adjust language to „French” aaY
aaSYS3 adjust language to „Spanish” aaY
aaVE Version number of the software aaVersion: