Yamatake Corporation Operation of the Transmitter
ST3000 Smart Transmitter Series 900 Electronic Differential Pressure/Pressure Transmitter 5-47
Step Description SFC screen
1
Press the key
The current output of the trans-
mitter will be displayed.
Change the displayed value to
50% by the following steps.
2
Press the and keys in
that order.
3
Press the key.
The SFC asks whether or not to
execute zero adjustment for 50%
output.
4
Press the key.
Zero adjustment has been com-
pleted.
OUT-
J
PUT
INPUT
OUTPUT FIT-1234
WORKING...
OUTPUT FIT-1234
45.00 %
S
5
Z
0
ACT PR
OUTPUT FIT-1234
50.00 %
G
SET
OUTPUT FIT-1234
SET LRV?
ENTER
( Yes )
NON-VOL
OUTPUT FIT-1234
50.00 %

Table of Contents