Muffle Furnace
FP102/302/312/412
Instruction Manual
First Edition
Thank you for purchasing “Muffle Furnace FP
series” of Yamato Scientific Co., Ltd.
This product has not been designed for medical
applications. Use this as a laboratory drying
sterilizer only.
In order to use this Equipment properly, please
read this Instruction Manual and Warranty Card
thoroughly before use. Keep them in safe place
close to this Equipment so that you can refer to
them any time.
Warning: Please read the important warning
notes in this Manual carefully and
thoroughly, and get the good
understanding of their contents
before using this Equipment.
Yamato Scientific America Inc
Santa ClaraCA
Printed on recycled paper

Table of Contents

Questions and Answers

  • K
    karen51Aug 4, 2025
    Why is the Yamato FP102 temperature display unstable?
    • D
      David NormanAug 4, 2025
      The temperature display on your Yamato Furnace might be unstable due to several reasons: * Heavy fluctuations in ambient temperature. In this case, you should review the location of the furnace. * Failure in the supplied power (the required voltage is ±10%). If so, connect the furnace to an adequate power supply. * Controller failure. If this is the case, replace the controller. * Temperature sensor failure. If so, replace the temperature sensor. * The furnace may be affected by the samples.
  • T
    timothysmithAug 13, 2025
    What to do if my Yamato Furnace doesn't display anything on the screen?
    • C
      Christopher McgeeAug 13, 2025
      If your Yamato Furnace isn't displaying anything on either the top or bottom screen when the controller power key is pressed and held, it could be due to: * Failure in the supplied power (the required voltage is ±10%). Connect the furnace to an adequate power supply. * Controller failure. If so, replace the controller.
  • J
    Jeffrey ThompsonAug 23, 2025
    How to fix the Yamato FP102 when the current time is not displayed?
    • A
      Angela MillerAug 25, 2025
      If the current time is not displayed on the bottom screen of your Yamato Furnace when the Earth Leakage Breaker (ELB) is ON, it could be due to: * No power supply. Check the connection to the power supply and apply power. * A failed ELB. Replace the ELB. * Controller failure. Replace the controller.

Summary

Safety Precautions

Symbols and Risk Information

Explains warning symbols, their meanings, and lists residual risks.

Before Operating the Equipment

Installation Precautions

Details site selection, leveling, ventilation, power, and grounding requirements.

Operating Procedure

Precautions on Handling

Handling Warnings and Cautions

Provides essential warnings for safe handling, operation, and material processing.

Maintenance Method

When a Trouble Occurs

Error Codes and Troubleshooting

Lists error codes, causes, solutions, and general troubleshooting guidance.

Specifications

Yamato FP102 Specifications

General IconGeneral
BrandYamato
ModelFP102
CategoryFurnace
LanguageEnglish