60
10. Specifications
Specifications
Product Name
Muffle Furnace
Model Name
FP102
FP302
FP312
FP412
Oroduct Code
214124
214128
214126
214127
System
Natural convection
Operating environment
temperature range
5℃~35℃
Power supply
Single phase AC115V
Single phase AC220V
Common to 50/60Hz, operating voltage range:±10%
Performance
※
1
Temperature
Control Range
100~1,150℃
Temperature
control precision※
2
±1.0℃ (at1150℃)
Temperature
fluctuation ※2
±1.0℃ (at1150℃)
Temperature
distribution
precision※2
±4.0℃ (at1150℃)
Temperature slope
※2
14.0℃ (at1150℃)
Temperature rise
time
Approx. 90 min.
Approx. 80 min.
Composition
Exterior
Chrome-free electro-galvanized steel plate Chemical proof baking finish
Furnace body
Ceramic fiber
Sensor
R thermo-couple
Heater
Iron chrome wire (Pyromax)
1.1kw
2.4kw
3.25kw
Exhaust port
I.D.20mm(Upper part)
Cooling fan
19/16W(50/60Hz)
Controller
Type
V-shaped
Temperature
Control Method
PID Z control
Temperature setting
method
Digital setting with ▲/▼ keys.
Temperature
Display Method
Setting temp. display:Orange 5-digit LED Digital Display (Resolution:1℃)
Temp. display:Green 4-digit LED Digital Display (Resolution:1℃)
Other displays
LED indicates temperature patterns for heating/stable/cooling
Timer/timer
resolution
Time: 1 minute and 99 hours 59 minutes/1min 24 hour setting
Operating function
Fixed temperature operation, quick auto stop, auto start, ,auto stop,
Program operation: Maximum 99 steps, up to 99 patterns, the repeat operation
function
Additional function
Power on and Operation Time Integrating Function(up to 65,535 hours);
calendar timer(24hours)
Calibration Offset; Power Consumption, Total CO2 Emission, and Heater
operating Output; Power Recovery Mode; Save and Access of Operater’s
Setting Information;、
Heater Control
Triac with Zero-cross Control
Temp/ sensor
R type Thermocouple x 2
(for temperature control and independent overheat preventive device)

Table of Contents