<7. Errors and Countermeasures (Display unit)>
63
IM 01E24A01-01EN
7. Errors and Countermeasures
(Display unit)
The error messages are described in the following tables.
For AXFA11, read the applicable user’s manual as listed in Table 1.1.
Explanation of NE107 status:
NE107 status Status of the device
F Failure Device malfunction, Parts malfunction
C Function Check The output signal is temporarily invalid for the local operation or manual operation.
S Out of Specication
The device works in out of specication.
The output signal is uncertain for the process or the ambience.
M Maintenance Required The maintenance is required in the near future.
N No Effect Other issue
System Alarm
Device breaks down and causes abnormal measurement.
Device replacement is needed.
NE107
Status
Error Message
Error Description Countermeasure
Display
F 010:Main CPU FAIL CPU (Main board) failure was detected. Contact Yokogawa service center.
F 011:Rev calc FAIL Failure of reverse calculation was detected. Contact Yokogawa service center.
F 012:Main EEP FAIL Failure of EEPROM (Main board) was detected.
Turn on the power again within the temperature range. If the problem
does not improve, contact Yokogawa service center.
F 013:Main EEP dt EEPROM (Main board) was reseted to default values. Contact Yokogawa service center.
F 014:Snsr bd FAIL Failure of sensor board was detected. Contact Yokogawa service center.
F 015:Snsr comm ERR Communication error of sensor was detected. Contact Yokogawa service center.
F 016:AD 1 FAIL[Sig]
Failure of A/D transumitter 1 [flow velocity signal]
wasdetected.
Contact Yokogawa service center.
F 017:AD 2 FAIL[Excit] Failure of A/D transmitter 2 [Exciting current] was detected. Contact Yokogawa service center.
F 018:Coil open Coil of sensor was disconnected. Turn off the power, check coil of sensor and excitation cable.
F 019:Coil short Coil of sensor was shorted. Contact Yokogawa service center.
F 020:Exciter FAIL Failure of excitation circuit was detected. Contact Yokogawa service center.
F 021:PWM 1 stop Error of pulse width modulation 1 was detected. Contact Yokogawa service center.
F 022:PWM 2 stop Error of pulse width modulation 2 was detected. Contact Yokogawa service center.
F 023:Opt bd mismatch Mismatch of option board was detected. Contact Yokogawa service center.
F 024:Opt bd EEP FAIL Failure of EEPROM (option board) was detected. Contact Yokogawa service center.
F 025:Opt bd A/D FAIL Failure of A/D (option board) was detected. Contact Yokogawa service center.
F 026:Opt bd SPI FAIL Failure of SPI (option board) was detected. Contact Yokogawa service center.
F 027:Restore FAIL Restore of parameters was failed. Retry parameter restoration.
F 028:Ind bd FAIL Failure of indicator board was detected.
Check the ambient temperature of display is within the range. If the
problem does not improve, contact Yokogawa service center.
F 029:Ind bd EEP FAIL Failure of EEPROM (indicator board) was detected.
Turn on the power again within the temperature range. If the problem
does not improve, contact Yokogawa service center.
F 030:LCD drv FAIL Failure of LCD driver was detected. Contact Yokogawa service center.
F 031:Ind bd mismatch Mismatch of Indicator board was detected. Contact Yokogawa service center.
F 032:Ind comm ERR Communication error of indicator board was detected. Check connection of Indicator & main board.
F 033:microSD FAIL Failure of microSD card was detected. Change microSD card.
F 244:Amp EEP FAIL Amplier EEPROM failed. Replace electrical parts such as the amplier. Or replace the device.
C 249:SoftDL Incmplt Software download is incomplete.
Check the cables, power supply and MTB SoftDL Error (MTB.
SOFTDL_ERROR)
C 250:SoftDL FAIL Software download has failed.
Check the download file and MTB SoftDL Error (MTB.SOFTDL_
ERROR)
F 251:Abnml Boot PRS Abnormal boot processing was detected at the startingtime.
Check the cables, power and MTB SoftDL Error (MTB.SOFTDL_
ERROR)
F 335:IT1 ttl bkup ERR
IT1 Total Backup has failed. Last IT1 Output.Value (IT1.
LAST.VALUE) is not saved.
Contact the nearest sales ofce or service center.
F 339:IT2 ttl bkup ERR
IT2 Total Backup has failed. Last IT2 Output.Value (IT2.
LAST.VALUE) is not saved.
Contact the nearest sales ofce or service center.
Errors and Countermeasures (Display unit)
7

Other manuals for YOKOGAWA AXW Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AXW Series and is the answer not in the manual?

YOKOGAWA AXW Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelAXW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals