Page #106 background imageLoading...
Page #106 background image
4-79
IM DL350-17EN
Commands
1
2
3
4
5
App
Index
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:GETMeasure
Function Executes the measurement of the linear scaling
P1:X or P2:X value of a 4-CH module, 16-CH
voltage input module, 16-CH temperature/voltage
input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:GETMeasure {P1X|P2X}
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:MODE
Function Sets or queries the linear scaling mode of a 4-CH
module, 16-CH voltage input module, or 16-CH
temperature/voltage input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:MODE {AXB|OFF|P12}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:MODE?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
Example
:CHANNEL1:VOLTAGE:SCHANNEL1:LSCALE:
MODE AXB
:CHANNEL1:VOLTAGE:SCHANNEL1:LSCALE:
MODE?
-> :CHANNEL1:VOLTAGE:SCHANNEL1:
LSCALE:MODE OFF
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:{P1X|P1Y|P2X|P2Y}
Function Sets or queries the linear scaling P1:X, P1:Y,
P2:X, or P2:Y value of a 4-CH module, 16-CH
voltage input module, or 16-CH temperature/
voltage input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:{P1X|P1Y|P2X|P2Y} {<NRf>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:{P1X|P1Y|P2X|P2Y}?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
<NRf>ofP1X,P2X=−9.9999E+30to−1E−30,0,
1E−30to9.9999E+30
<NRf>ofP1Y,P2Y=−9.9999E+30to−1E−30,0,
1E−30to9.9999E+30
Example
:CHANNEL1:VOLTAGE:SCHANNEL1:LSCALE:
P1X 1
:CHANNEL1:VOLTAGE:SCHANNEL1:LSCALE:
P1X?
-> :CHANNEL1:VOLTAGE:LSCALE:P1X 1
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:UNIT
Function Sets or queries the linear-scaling unit string of a
4-CH module, 16-CH voltage input module, or 16-
CH temperature/voltage input module.
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:UNIT {<String>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:UNIT?
4-CH module (720254)
<x1> = 1 to 4
<x2> = 1, 2
16-CH Voltage Input Module (720220)
<x1> = 1, 3
<x2> = 1 to 16
16-CH Temperature/Voltage Input Module
(720221)
<x1> = 1, 3
<x2> = 1 to 16
<String> = Up to 4 characters
Example
CHANNEL1:VOLTAGE:LSCALE:UNIT “UU”
CHANNEL1:VOLTAGE:LSCALE:UNIT?
-> CHANNEL1:VOLTAGE:LSCALE:UNIT “UU”
4.6 CHANnel Group

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals