4-112
IM DL350-17EN
:GONogo:PARameter:ITEM<x>:TYPE?
Function Queries, for parameter determination, the
specified waveform parameter’s measurement
item and upper and lower limits.
Syntax
:GONogo:PARameter:ITEM<x>:TYPE?
<x> = 1 to 8
Example
:GONOGO:PARAMETER:ITEM1:TYPE?
-> :GONOGO:PARAMETER:ITEM1:TYPE:
MAXIMUM 1.10000E+00,1.00000E+00
:GONogo:PARameter:ITEM<x>:TYPE:
<Parameter>
Function Sets or queries the upper and lower limits of the
measurement item for the specified waveform
parameter.
Syntax
:GONogo:PARameter:ITEM<x>:TYPE:
<Parameter> {<Voltage>,<Voltage>|
<Current>,<Current>|<Time>,<Time>|
<Frequency>,<Frequency>|<NRf>,<NRf>}
<x> = 1 to 8
<Parameter> = {AMPLitude|AVERage|
AVGFreq|AVGPeriod|BWIDth1|BWIDth2|
DELay|DUTYcycle|FALL|FREQuency|
HIGH|LOW|MAXimum|MIDDle|MINimum|
NOVershoot|NWIDth|PERiod|PNUMber|
POVershoot|PTOPeak|PWIDth|RISE|
RMS|SDEViation|TY1Integ|TY2Integ}
:GONogo:PARameter:ITEM<x>:VALue?
Function Queries the measured value of the specified
waveform parameter.
Syntax
:GONogo:PARameter:ITEM<x>:VALue?
<x> = 1 to 8
:GONogo:REMote
Function Sets or queries the remote mode of GO/NO-GO
determination.
Syntax
:GONogo:REMote {<Boolean>}
:GONogo:REMote?
Example
:GONOGO:REMOTE ON
:GONOGO:REMOTE? -> :GONOGO:REMOTE 1
:GONogo:SEQuence
Function Sets or queries the action mode of GO/NO-GO
determination.
Syntax
:GONogo:SEQuence {CONTinue|SINGle}
:GONogo:SEQuence?
Example
:GONOGO:SEQUENCE CONTINUE
:GONOGO:SEQUENCE?
-> :GONOGO:SEQUENCE CONTINUE
:GONogo:TRANge
Function Sets or queries the determination range of GO/
NO-GO determination.
Syntax
:GONogo:TRANge {<NRf>,<NRf>}
:GONogo:TRANge?
<NRf> = –5 to 5 div
Example
:GONOGO:TRANGE−5,5
:GONOGO:TRANGE?
->:GONOGO:TRANGE−5.000000000000,
5.000000000000
:GONogo:WAIT?
Function Waits for the completion of GO/NO-GO
determination with a timeout period.
Syntax
:GONogo:WAIT? {<NRf>}
<NRf> = 1 to 864000 (in units of 100 ms)
Example
Setting the timeout period to 5 seconds
:GONOGO:WAIT?50->0
Description The instrument returns 0 if the operation finishes
before the timer expires and returns 1 if the timer
expires.
:GONogo:ZONE?
Function Queries all waveform zone determination
settings.
Syntax
:GONogo:ZONE?
:GONogo:ZONE:PATTern<x>?
Function Queries all settings for the specified determination
pattern for waveform zone determination.
Syntax
:GONogo:ZONE:PATTern<x>?
:GONogo:ZONE:PATTern<x>:CAUSe?
Function Queries whether the specified determination
pattern for waveform zone determination is the
cause of a NO-GO judgment.
Syntax
:GONogo:ZONE:PATTern<x>:CAUSe?
<x> = 1 to 8
Example
:GONOGO:ZONE:PATTERN1 CAUSE?
-> :GONOGO:ZONE:PATTERN1 CAUSE 0
Description When the determination pattern is the cause
of a NO-GO result, the instrument returns 1.
Otherwise, the instrument returns 0.
:GONogo:ZONE:PATTern<x>:MODE
Function Sets or queries the specified determination
pattern’s reference condition for waveform zone
determination.
Syntax
:GONogo:ZONE:PATTern<x>:MODE {IN|
OUT|OFF}
:GONogo:ZONE:PATTern<x>:MODE?
<x> = 1 to 8
Example
:GONOGO:ZONE:PATTERN1:MODE IN
:GONOGO:ZONE:PATTERN1:MODE?
-> :GONOGO:ZONE:PATTERN1:MODE IN
4.14 GONogo Group (Scope mode only)