4-131
IM DL350-17EN
Commands
4
:MEASure:{CHANnel<x1>[:SCHannel<x2>]
|MATH<x3>}:<Parameter>?
Function Queries the setting of a waveform parameter
(measurement item).
Syntax
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:<Parameter>?
<x1> = 1 to 4, 7
<x2> = 1 to 60 (sub channel)
<x3> = 1, 2
<Parameter> = {AMPLitude|AVERage|
AVGFreq|AVGPeriod|BWIDth1|BWIDth2|
DUTYcycle|FALL|FREQuency|HIGH|LOW|
MAXimum|MIDDle|MINimum|NOVershoot|
NWIDth|PERiod|PNUMber|POVershoot|
PTOPeak|PWIDth|RISE|RMS|
SDEViation|TY1Integ|TY2Integ|}
Example
:MEASURE:CHANNEL1:PTOPEAK?
-> :MEASURE:CHANNEL1:PTOPEAK:STATE 0
Description • For channels that have sub channels, use the
:MEASure:CHANnel:SCHannel:<Parameter>?
command.
• This command returns the same response as
the :MEASure:CHANnel:<Parameter>:STATe?
command.
:MEASure:{CHANnel<x1>[:SCHannel<x2>]
|MATH<x3>}:<Parameter>:COUNt?
Function Queries the count of measured values for cyclic
statistical processing of a waveform parameter.
Syntax
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:<Parameter>:COUNt?
<x1> = 1 to 4, 7
<x2> = 1 to 60 (sub channel)
<x3> = 1, 2
<Parameter> = {AMPLitude|AVERage|
AVGFreq|AVGPeriod|BWIDth1|BWIDth2|
DUTYcycle|FALL|FREQuency|HIGH|LOW|
MAXimum|MIDDle|MINimum|NOVershoot|
NWIDth|PERiod|PNUMber|POVershoot|
PTOPeak|PWIDth|RISE|RMS|
SDEViation|TY1Integ|TY2Integ}
Example
:MEASURE:CHANNEL1:PTOPEAK:COUNT?
-> :MEASURE:CHANNEL1:PTOPEAK:COUNT 3
Description For channels that have sub channels, use the
:MEASure:CHANnel:SCHannel:<Parameter>:CO
UNt? command.
:MEASure:{CHANnel<x1>[:SCHannel<x2>]
|MATH<x3>}:<Parameter>:{MAXimum|MEAN
|MINimum|SDEViation}?
Function Queries a cyclic statistical processing value of a
waveform parameter.
Syntax
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:<Parameter>:{MAXimum|MEAN|
MINimum|SDEViation}?
<x1> = 1 to 4, 7
<x2> = 1 to 60 (sub channel)
<x3> = 1, 2
<Parameter> = {AMPLitude|AVERage|
AVGFreq|AVGPeriod|BWIDth1|BWIDth2|
DUTYcycle|FALL|FREQuency|HIGH|LOW|
MAXimum|MIDDle|MINimum|NOVershoot|
NWIDth|PERiod|PNUMber|POVershoot|
PTOPeak|PWIDth|RISE|RMS|
SDEViation|TY1Integ|TY2Integ}
Example
:MEASURE:CHANNEL1:PTOPEAK:MAXIMUM?
-> :MEASURE:CHANNEL1:PTOPEAK:
MAXIMUM 30.633333E+00
Description • For channels that have sub channels, use the
:MEASure:CHANnel:SCHannel:<Parameter
>:{MAXimum|MEAN|MINimum|SDEViation}?
command.
• This command returns “NAN” for any statistics
that cannot be processed.
:MEASure:{CHANnel<x1>[:SCHannel<x2>]
|MATH<x3>}:<Parameter>:STATe
Function Sets or queries whether the specified waveform’s
waveform parameter (measurement item) is ON
or OFF.
Syntax
:MEASure:{CHANnel<x1>[:SCHannel<x2>]|
MATH<x3>}:<Parameter>:STATe
<x1> = 1 to 4, 7
<x2> = 1 to 60 (sub channel)
<x3> = 1, 2
<Parameter> = {AMPLitude|AVERage|
AVGFreq|AVGPeriod|BWIDth1|BWIDth2|
DUTYcycle|FALL|FREQuency|HIGH|LOW|
MAXimum|MIDDle|MINimum|NOVershoot|
NWIDth|PERiod|PNUMber|POVershoot|
PTOPeak|PWIDth|RISE|RMS|
SDEViation|TY1Integ|TY2Integ}
Example
:MEASURE:CHANNEL1:PTOPEAK:STATE ON
:MEASURE:CHANNEL1:PTOPEAK:STATE?
-> :MEASURE:CHANNEL1:PTOPEAK:STATE 1
Description For channels that have sub channels, use the
:MEASure:CHANnel:SCHannel:<Parameter>:ST
ATe command.
4.22 MEASure Group