4-150
IM DL350-17EN
:SYSTem:CLOCk:SNTP?
Function Queries all SNTP settings.
Syntax
:SYSTem:CLOCk:SNTP?
:SYSTem:CLOCk:SNTP:EXECute
Function Uses SNTP to set the date and time.
Syntax
:SYSTem:CLOCk:SNTP:EXECute
Example
:SYSTEM:CLOCK:SNTP:EXECUTE
:SYSTem:CLOCk:SNTP:GMTTime
Function Sets or queries the time difference from GMT
when SNTP is being used.
Syntax
:SYSTem:CLOCk:SNTP:
GMTTime {<String>}
:SYSTem:CLOCk: SNTP:GMTTime?
<String>=HH:MM(−12:00to13:00)
Example
:SYSTEM:CLOCK:SNTP:GMTTIME "09:00"
:SYSTEM:CLOCK:SNTP:GMTTIME?
-> :SYSTEM:CLOCK:SNTP:
GMTTIME "09:00"
:SYSTem:CLOCk:TIME
Function Sets or queries the time.
Syntax
:SYSTem:CLOCk:TIME <String>
:SYSTem:CLOCk:TIME?
<String> = HH:MM:SS
HH can be set to a value from 0 to 23.
Example
:SYSTEM:CLOCK:TIME "10:00:00"
:SYSTEM:CLOCK:TIME?
-> :SYSTEM:CLOCK:TIME "10:00:00"
:SYSTem:CLOCk:TYPE
Function Sets or queries whether the date and time are set
manually or by using SNTP.
Syntax
:SYSTem:CLOCk:TYPE {MANual|SNTP}
:SYSTem:CLOCk:TYPE?
Example
:SYSTEM:CLOCK:TYPE MANUAL
:SYSTEM:CLOCK:TYPE?
-> :SYSTEM:CLOCK:TYPE MANUAL
:SYSTem:CRMode (Cursor Read Mode)
Function Sets or queries the mode used to read vertical,
marker, and degree cursor values.
Syntax
:SYSTem:CRMode {DISPlay|ACQuisition}
:SYSTem:CRMode?
Example
:SYSTEM:CRMODE DISPLAY
:SYSTEM:CRMODE?
-> :SYSTEM:CRMODE DISPLAY
:SYSTem:HAFormat(Horizontal Axis
Format)
Function Sets or queries the horizontal axis display format.
Syntax
:SYSTem:HAFormat {AUTO|ABSolute|
RELative}
:SYSTem:HAFormat?
AUTO: Auto
ABSolute: Absolute time
RELative: Relative time
Example
:SYSTEM:HAFORMAT AUTO
:SYSTEM:HAFORMAT?
-> :SYSTEM:HAFORMAT AUTO
:SYSTem:KEYProtect?
Function Queries all key and touch lock settings.
Syntax
:SYSTem:KEYProtect?
:SYSTem:KEYProtect:EPASsword (Enter
password)
Function Enters the password for releasing the main unit’s
key and touch lock.
Syntax
:SYSTem:KEYProtect:
EPASsword {<String>}
<String> = Up to 8 characters
Example
:SYSTEM:KEYPROTECT:EPASSWORD "ABC"
Description This command is valid when the keys and touch
panel are locked, a password is required to
release the key and touch lock, and a password
has been set.
:SYSTem:KEYProtect:MODE
Function Sets or queries whether the main unit’s keys and
touch panel are locked.
Syntax
:SYSTem:KEYProtect:MODE {<Boolean>}
:SYSTem:KEYProtect:MODE?
Example
:SYSTEM:KEYPROTECT:MODE ON
:SYSTEM:KEYPROTECT:MODE?
-> :SYSTEM:KEYPROTECT:MODE 1
Description The key and touch lock status is retained even
when the instrument is restarted.
4.31 SYSTem Group