<2. Handling Cautions>
2-16
IM 01C25F01-01E
Model
Capsule
code
PS*
1
(bar)
V(L)
PS.V
(bar.L)
Category*
2
EJ510
EJX610A
A, B, C 100 0.1 10
Article 4,
Paragraph 3
(SEP)
D 700 0.1 70
EJ510,
EJX610A
with code /
PE3
D 700 0.1 70 III
EJ530,
EJX630A
A, B, C 100 0.1 10
Article 4,
Paragraph 3
(SEP)
D 700 0.1 70
EJ530,
EJX630A
with code /
PE3
D 700 0.1 70 III
*1: PS is maximum pressure for vessel itself based on
Pressure Equipment Directive 2014/68/EU. Refer to
GeneralSpecicationformaximumworkingpressureofa
transmitter.
*2: Referred to Table 1 covered by ANNEX II of EC Directive
on Pressure Equipment Directive 2014/68/EU.
(3) Operation
CAUTION
• Thetemperatureandpressureofuidshould
be maintained at levels that are consistent
with normal operating conditions.
The ambient temperature should be
maintained at a level that is consistent with
normal operating conditions.
Please take care to prevent water hammer
and the like from inducing excessive
pressures in pipes and valves. If phenomena
are likely, install a safety valve or take
some other appropriate measure to prevent
pressure from exceeding PS.
Take appropriate measures at the device or
system level to protect transmitters if they
are to be operated near an external heat
source.
2.12 EU RoHS Directive
Applicable standard: EN 50581
Applicable production sites are shown below.
The condition of the RoHS compliant production
sites are as follows:
Japan, USA, Germany, Bahrain, India
Theproductionsitescanbeconrmedbythe
serial number shown in the frame of “NO.” in
the name plate of the product.
Serial numbers (9 letters): AAnnnnnnn
AA:Identicationcodeofproductionsite
Japan: Use “91” USA: Use “U1”
Germany: Use “D1” Bahrain: Use “BH”
India: Use “Y1”
2.13 Safety Requirement
Standards
Applicable standard: EN 61010-1, C22.2 No.61010-1
(1) Pollution Degree 2
“Pollution degree” describes the degree to
which a solid, liquid, or gas which deteriorates
dielectric strength or surface resistivity is
adhering. “ 2 ” applies to normal indoor
atmosphere. Normally, only non-conductive
pollution occurs. Occasionally, however,
temporary conductivity caused by condensation
must be expected.
(2) Installation Category I
“Overvoltage category (Installation category)”
describesanumberwhichdenesatransient
overvoltage condition. It implies the regulattion
for impulse withstand voltage. “ I ” applies to
electrical equipment which is supplied from the
circuit when appropriate transient overvoltage
control means (interfaces) are provided.
(3) Indoor/Outdoor use