<7. Operation>
7-9
IM 01C25F01-01E
7.5.11 Save or Cancel
At the end of each parameter setting, select “Save”
or “Cancel” by the external adjustment screw
and press the push button to save or cancel the
conguration.
Save? Saved
Save or Cancel
Cancel? Canceled
screw
F0713.ai
push
push
After
parameter
setting
7.5.12 AbortConguration
7.5.12.1AbortConguration(Menu)
Hold down the push button for over 2 seconds to
exit the Local Parameter Setting mode.
Hold down the button
for over 2 seconds
Proccess Value
Process Measurement Display
Process Measurement Display
1. Tag Number
2. Press Unit
3. Press LRV
4. Press URV
5. Damping Time
6. Output Mode
7. Display Out1
8. LRV-Apply Press
9. URV-Apply Press
F0714.ai
Jump to the normal mode
7.5.12.2AbortConguration(Parameter)
Toexitthecongurationwhileeditingthevalue,
hold down the button for over 2 seconds and select
“Save” or “Cancel”.
F0715.ai
Return to
parameter review window
Save or Cancel?
Push for
Over 2 seconds
Tag editing
(3
rd
Character)
7.5.13 Local Parameter Setting Lock
To disable parameter changes by the Local
Parameter Setting there are three different ways.
Locked features
Communication Parameter
Ext SW =disable
(EXT ZERO ADJ = disable)
External Zero
Adjustment
Local Parameter Setting
Communication Parameter
Write Protect = On
(WRT PROTECT = Yes)
Local Parameter Setting
All Communication
Parameters *
Hardware write protection
switch on CPU assembly =
D (Disable)
Local Parameter Setting
All Communication
Parameters *
*External Zero Adjustment is unlocked.
The above parameter setting is carried out by
usingeldcommunicatororDTM.Seetheusers
manual IM 01C25T (HART/BRAIN) for the setting
procedure.
Reviewing local parameter setting by push button
on the integral indicator is available at any time
even when the Local Parameter Setting is locked.
7.5.14 Others
Difference between BRAIN and HART
HART BRAIN
Tag Number
Up to 8
characters can
be set
Up to 16
characters can
be set
Output Mode
(Linear/Sq root)
“TRNS.FC”
shown on the
integral indicator
“OUT.MOD”
shown on the
integral indicator
The degree of adjustment depends on the
speed of turning the adjustment screw. Turn the
screwslowlyfornetuningandturnthescrew
fast for quick tuning.