5-16
IM 04L51B01-01EN
Code Message Description, Corrective Action, Ref. Section
281 Exceeds time deviation setting. The time change made during recording exceeds
the time deviation limit. Set the time within the Time
deviation limit, which is set in Time basic settings
of System settings.
page 1-168 in section 1.19.4,
“Setting the Time Zone, Gradual Time Adjustment,
and Daylight Saving Time”
301 No modules to download the firmware. Do not remove the module while firmware is being
downloaded.
302 Firmware file not found. Check that the firmware file exists on the specified
medium.
303 Failed to read firmware file. Download the firmware again. Check that the
SD memory card is not damaged and that it is
not disconnected while the firmware is being
downloaded.
304 Corruputed firmware file. Download the appropriate firmware from our Website.
305 Invalid firmware version. Download the appropriate firmware from our Website.
306 Mismatch module type. Download the appropriate firmware for the module
type that you want to update from our Website.
307 Firmware download error. Download the firmware again. Check that the
SD memory card is not damaged and that it is
not disconnected while the firmware is being
downloaded.
•OperationErrors
Code Message Description, Corrective Action, Ref. Section
501 Operation aborted due to media error. Change the storage medium.
502 Not enough free space on media. Not enough free space on the storage medium, or the
limit of the number of directories has been exceeded.
Change the storage medium.
503 Media not recognized. Remove the storage medium, and set it again.
504 Media read error. Change the storage medium.
505 Media write error.
506 Failed to create the file.
507 No file or directory. Tried to access a file which does not exists in the
internal memory.
508 Format error. Reformat.
509 Unknown file type. Check the extension.
511 Invalid file or directory operation. Tried to delete a directory containing files. Delete
the files and directories in the directory first before
executing the operation.
512 The file is already in use. Try again later. Wait until the file is accessible.
514 No setting files. Check the setting parameter file.
515 A template file is abnormal. Check the report template file.
516 Some items not set. Check the error log.
517 Media is read-only. Make it writable.
518 Scale image not loaded. Load a scale image.
519 Template file not loaded. Load a template file.
532 No data available. This error occurs when there is no valid data in the
internal memory. Specify valid data.
536 No data after the date and time. Select an earlier time.
537 The specified screen cannot be displayed Specify a valid screen.
539 No selected data. This error occurs when there is no valid data in the
internal memory. Specify valid data.
541 Printer not connected. Check the printer settings. Check the printer and
network conditions.
542 Printer busy. Execute after printing is complete.
551 Module was detached. Check that the module is installed correctly.
552 Detected newer version module. Check the module version.
553 Unrecognized module. Check that the module is installed correctly.
If the same message continues to appear even
after you perform the procedure above, servicing is
required. Contact your nearest YOKOGAWA dealer
for repairs.
554 Fatal module error.
555 Non-calibrated module. Perform A/D calibration.
556 RJC error module. Servicing is required. Contact your nearest
YOKOGAWA dealer for repairs.
557 A/D error module.
558 Activation required for attached modules. Activate modules.
Continued on next page
5.2 Troubleshooting