3-8 Installation wiring
Figure 3-9a. Impedance measurement jumper setting
2
1
Default jumper connections for :
- Input 1, High impedance
- Input 2, Low impedance
Service code 03 = 1.1.1
- Input 1, High impedance, Temp. comp., Check on
Service code 04 = 0.0.1
- Input 2, Low impedance, Temp. comp. off, Check on
Jumper connections for :
- Input 1, High impedance
- Input 2, High impedance
Service code 03 = 1.0.1
- Input 1, High impedance, Temp. comp. off, Check on
Service code 04 = 1.0.1
- Input 2, Low impedance, Temp. comp. off, Check on