Operation manual of ZBL-R800, ZBL-R630, ZBL-C310A
Hotline: (86)010-51290405 Fax: (86)010-51290406
54
designed value.
Operation steps:
1) To set default diameter of tested rebar diameter.
2) To set the minimum cover thickness value for alarm.
3) While scanning the tested object, when thickness of cover is
less than the set value, the buzzer alarms.
Under this measurement mode, users can scan in a relative fast speed
without noticing screen display.
3.4.6.6 Rebar diameter measurement
Table 3.2 Measuring range of rebar diameter (unit: mm)
Rebar diameter Minimum thickness Maximum thickness
Ф6
4 55
Ф8
4 60
Ф10
6 62
Ф12
8 64
Ф14
10 66
Ф16
10 68
Ф18
12 68
Ф20
12 70
Ф22
14 72
Ф25
14 74
Ф28
16 76
Ф32
16 77
Ф36
16 78
Ф40
20
80
Ф50
22 84
Firstly, to locate rebar accurately and to identify precise direction of
rebar, placing sensor above the tested rebar and pressinga key, the
word “Dia.” will be displayed on the screen. After a few minutes,
diameter measurement will be shown directly on the screen and the
tested thickness value of cover shown after “Cover”. “”is before
this value to distinguish from thickness value tested by default