General Safety Information
MCS 600
2.7 Storage and transport
No special transport container or locking device is provided with the system.
Although the components are very robust, collision of any sort must be avoided.
Install the measuring system in a safe, easily accessible location. Please note the
specified environmental parameters when storing and working with the system.
2.8 Installation and connection requirements
The MCS 600 system is designed to be used in a horizontal position with the
housing feet facing downward. We do not recommend placing the device in any
other position, as this will impair the device's ability to dissipate heat. This also
applies to installations involving a protective housing (an optional accessory for
using the MCS 600 in environments requiring higher IP ratings).
Line voltage 100 – 240 V AC
Power frequency 50/60 Hz
Working temperature range
*exception for MCS 611 NIR 1.7 for lab applications
(000000-2187-068)
+5 °C to 35 °C*
+5 °C to 25 °C
Storage/transport temperature – 40 °C to + 70 °C
Max. humidity 95 % non-condensing
Use environment altitude above sea level Up to 2000 m