MCS 600
3.2.5 IN/OUT configuration
Fig. 3 19-inch housing,
IN/OUT configuration
IN 0
IN 1
OUT 7
NC
NC
Aux 24 V
IN 0
IN 1
IN 2
IN 3
IN 4
IN 5
IN 6
IN 6
OUT 0
OUT 1
OUT 2
OUT 3
OUT 4
OUT 5
OUT 6
OUT 7
Aux 24 V
Aux 24 V
Aux 24 V
Aux 24 V
GND
GND
GND
GND
IN 0 to 3 Aux 24 V
OUT 0GND
S1 Relay
GND
GND
1
1
17
16
4
1
Trigger IN Trigger OUT
32
8
5
GND
GND
GND
NC
NC
GND
Note
In this version, only IN 0 through IN 5 and OUT 0 through OUT 7 can be
used.
Warning
Maximum load: 24 V DC: 800 mA
Maximum load: OUT 0 through 7: 100 mA each
Only connect Trigger IN to bounce-free and potential-free contacts.
Overload may cause the system to reset!