8.2 Physiological Data Anomalies
Excessively
high or
erratic
Movement of
sensor surface
against skin
Check strap tension – has it loosened?
Is skin excessively dry
Constant
compression of
breathing sensor
by external force
Check for / remove external agent
Check device at apex of rib curvature
Exit OmniSense, disconnect the
Gateway receiver, wait & restart
OmniSense
Check antenna connections
Confirm no duplication of device
Short Addresses, especially if a new
device has been added to the system.
Use Zephyr Config Tool
Device internal
clock not
resynchronized
Connect device to PC and start log
import utility in OmniSense Analysis
to reset internal clock
Use Zephyr Cfg Tool to resync time
8.4 General Check Procedure
Carry out the following checks when in Live mode
Check LED behavior Section 2.3.2