Page #49 background imageLoading...
Page #49 background image
4948
Appendices
SW Update
ID
C
T
Full text
Corrective action
E5XX
E
U
Missing source: ……
Check and verify the software
update sources on the media.
Restart update.
E5XX
E
U
Source check failure: ……
Corrupted or missing files. Check
and verify the software update
sources on the media. Restart
update.
E5XX
E
U
Unpack failed: ……
Corrupted or missing files. Check
and verify the software update
sources on the media. Restart
update.
I5XX
I
U
……
N/A
O5XX
I
U
……
N/A
13 Appendices
Appendix A Results table
The DocUReader 2 Pro analyzer prints the results in the following gradation of
concentration using LabStrip
urine test strips:
Parameter Conventional Units
(Conv.)
SI Units (SI) Arbitrary
BIL
(Bilirubin)
negative
0.5 mg/dl
1 mg/dl
3 mg/dl
6 mg/dl
negative
8.5 µmol/l
17 µmol/l
50 µmol/l
100 µmol/l
negative
UBG
(Urobilinogen)
normal
2 mg/dl
4 mg/dl
8 mg/dl
12 mg/dl
normal
35 µmol/l
70 µmol/l
140 µmol/l
200 µmol/l
negative
1+
2+
3+
4+
KET
(Ketone)
negative
5 mg/dl
15 mg/dl
50 mg/dl
150 mg/dl
negative
0.5 mmol/l
1.5 mmol/l
5 mmol/l
15 mmol/l
negative
(+)
1+
2+
3+
ASC
(Ascorbic acid)
negative
20 mg/dl
40 mg/dl
100 mg/dl
negative
20 mg/dl
40 mg/dl
100 mg/dl
negative
1+
2+
3+