Maintenance
11 Maintenance
11.1 Cleaning the analyzer
It is recommended to keep the DocUReader 2 Pro device clean and dust
free.
Always make sure that the analyzer is switched off before cleaning.
Do not turn the analyzer on its side or upside down during cleaning as
previously spilled urine or cleaning liquid could run inside the case and
damage electrical parts.
Make sure that no liquid enters the device and printer compartment.
Do not use any type of solvent, oil, grease, silicone spray or lubricant on
the analyzer.
Don’t use any sprayer/atomizer for cleaning the device! Use only wet
towel dipped in a mild detergent.
Recommended cleaning agents:
• Isorapid (a mixture of 20 g Ethanol, 28 g 1-Propanol, and 0.1 g quaternary
ammonium compounds)
• Trigene Advance laboratory disinfectant (at a dilution of 1:100)
• Barrycidal-33 (at a dilution of 2:100)