97
COMMAND
NAME
DESCRIPTION No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
Status LED Read and set current Status LED mode:
0 – Normal
1 – F. Alarm & Totalizers only
2 – Alarm Events only
3 – Diagnostic Events
4 – UART interface events
5 – ModBus interface
19 S No Argument
(Returns Current
Status LED mode)
S:<mode value>
Example:
S:0
<mode value> [0-5]
Set new Status LED
mode value
S:< mode value >
Example:
S:1
Mass Flow
Units of
measure
Set units of measure for mass
fl ow rate and totalizer reading.
Note: The units of the totalizer
output are mass fl ow units
and not per unit time.
For users defi ned units:
k-Factor value represents conversion
value to L/min.
Time base argument:
0 - Seconds
1 - Minutes
2 - Hours
3 - Days
Density argument:
1 - Use density
0 - Do not use density
20 U
%FS* U: %FS
SuL/min U: SuL/min
Sml/sec U: SmL/sec
Sml/min U: SmL/min
Sml/hr U: SmL/hr
SL/sec U: SL/sec
SL/min U: SL/min
SL/hr U: SL/hr
SL/day U: SL/day
Sm3/min U: Sm3/min
Sm3/hr U: Sm3/hr
Sm3/day U: Sm3/day
Sf3/sec U:Sf3/sec
Sf3/min U:Sf3/min
Sf3/hr U:Sf3/hr
Sf3/day U:Sf3/day
gr/sec U: gr/sec
gr/min U: gr/min
gr/hr U: gr/hr
gr/day U: gr/day
kg/min U: kg/min
kg/hr U: kg/hr
ay U: kg/day
lb/min U: lb/min