102
COMMAND
NAME
DESCRIPTION No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
A
OLED Flow Running
Average [0-25]
<new value>
[0-25] samples
LA:<value>
Example: LA:1
No Argument
(Returns Current settings)
LA:<value>
Example: LA:1
E
OLED Flow reading Dead
Band in % F.S. [0.0 - 0.99]
<new value>
[0.0-0.99] %FS
LE:<value>
Example: LE:0.01
No Argument
(Returns Current Settings)
LE:<value>
Example: LE:0.01
No Argument Returns
OLED support status: Y or N
L:Y
Auto Zero Sets/Reads Instrument Auto Zero
related parameters
WARNING: make sure absolutely
no fl ow through the instr. during
SensorZero offset calibration!
NOTE: For proper result instr.
has to be connected to power for
at least 15 minutes prior to Auto
Zero calibration.
Auto Zero Status return parameters:
<T value> - Current Tare value
<ADC value> - Current ADC value
AZ Status:
N – Auto Zero Not Started
I – Auto Zero In Process
F – Auto Zero Failed
D – Auto Zero is Done (Success)
WARNING: make sure instr. is
open to atmosphere and absolutely
no fl ow through the instr. during
AP Sensor tare procedure! New AP
Value must be taken from
reference AP standard in PSIA!
25 Z
V
Display current Zero Value
No Argument ZV:< Value>
Example: ZV,589
N
Start Sensor Auto Zero
calibration now.
NOTE: make sure absolutely
no fl ow through the instrument!
No Argument ZN
NOTE: For proper result
instr must be connected
to power for at least
15 minuetes prior to Auto
Zero Calibration.
S
Display Flow Auto Zero Status
No Argument ZS: <T value> ADC value>
<AZ Status
Ex:ZS:7492,70581,N
A
Absolute Pressure Sensor
Tare request and status
NOTE: make sure absolutely
no fl ow through the instr.!
New AP Value must be taken
from reference AP standard!
No Argument
(Returns Current settings
and status)
APZ:<OffsetValue>,<ADC
value>,<TempCounts,
<AZstatus>,
Example: APZ:-00266
640002,839,N
<New Value>
[13.1 – 15.99] PSIA
ZA: <value>
Example: ZA
B
Background Auto Tare
No Argument
(Returns Current Set.)
AB:<value>
AB:0
<New Value>[0-1]
0-Disabled,1-Enabled
AB:<value>
AB:0