87
COMMAND
NAME
DESCRIPTION No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
Diagnostic
Events
Registers
Read/Set/Reset current value of the Diagnostic
Events Registers
Status Register: (Read/Reset)
Mask Register: (R/W)
Latch Register: (R/W)
See list of the Diagnostic Events below:
0 CPU_TEMP_HIGH 0x0001
1 DP EE INIT ERROR 0x0002
2 AP EE INIT ERROR 0x0004
3 VREF_OUT_OF_RANGE 0x0008
4 FLOW ABOVE LIMIT 0x0010
5 AP OUT OF RANGE 0x0020
6 G TEMP OUT OF RANGE 0x0040
7 ANALOG OUT ALARM 0x0080
8 SER COMM FAILURE 0x0100
9 MB COMM FAILURE 0x0200
A EEPROM FAILURE 0x0400
B AUTOZERO FAILURE 0x0800
C AP TARE FAILURE 0x1000
D DP PRESSURE INVALID 0x2000
E AP PRESSURE INVALID 0x4000
F FATAL_ERROR 0x8000
6 DE NO ARGUMENT
(read Diagnostic
Events status
register)
DE:<Value>
Example: DE:0x0
R
(reset Diagnostic
Events status
register
to 0x0000)
DER:0x0
M
(Read/Set
Diagnostic
Events Mask
register)
NO ARGUMENT
(read current
settings)
DEM:0x1
<Value>
0x0000-0x0FFF
Set new value
NOTE: all 6
characters are
required
DEM:0x101
L
(Read/Set
Diagnostic
Events Latch
register)
NO ARGUMENT
(read current
settings)
DEL:0x1
<Value>
0x0000-0x0FFF
Set new value
NOTE: all 6
characters are
required
DEL:0x101