88
COMMAND
NAME
DESCRIPTION No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
Read Gas
Temperature
Requests the Gas Temperature
reading in current TEU
7 GT NO ARGUMENT
(read only)
<T Value> (Actual gas
temperature in current
temp. engineering units)
Example: 24.51
Read Gas
Pressure
Requests the Gas Pressure reading
in current PEU
8 GP NO ARGUMENT
(read only)
<P Value> (Actual gas
pressure in current
pressure engineering units)
Example: 14.66
Gas Read / Select Active Gas Indexes:
NOTE: Instrument confi gured for non-corrosive
gases support indexes 0 to 128 (see list of all
supported gases).
9 G NO ARGUMENT
(read current active Gas
index and Gas Name)
Example: G:0,AIR
0 – Gas Index
AIR – Gas name
<Value>
[0-128]
Select new Gas
NOTE: Instruments
w/o Corrosive Gases
support only 129 gases
[0-128]
Example: G:5,He
5 – Gas Index
He – Gas name