COMMAND
NAME
DESCRIPTION No.
COMMAND SYNTAX
Command Argument 1 Argument 2 Argument 3 Argument 4 Response
Device Info Read device confi guration info:
- Currently selected Gas (index, name)
- Full scale range (L/min)
- Mass fl ow Units of measure
- Volumetric fl ow Units of Measure
- Totalizer#1 mode
D – Disabled
E – Enabled
- Totalizer#2 mode
D – Disabled
E – Enabled
- Analog Output Mode
0 – 0-5 Vdc
1 – 0-10 Vdc
2 – 4-20 mA
- ModBus H/V status [0/1]
0 – Installed
1 – Not Installed
10 DI NO ARGUMENT
(Read Only)
DI:5,Helium,0.200,
Sml/min,ml/min,E,D,0,1
5 – Gas index 5(Helium)
0.200 – full scale (L/min)
Sml/min – current MEU
ml/min – current VEU
E – Totalizer#1 Enabled
D – Totalizer#2 Disabled
0 – Analog Output set to 0-5 Vdc
1 – ModBus interface H/W is
not installed (Not Supported)