EasyManua.ls Logo

Abaxis VetScan HM5 - VetScan HM5 Specifications

Abaxis VetScan HM5
164 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10-2 Specifications
10.1 VetScan HM5 Specifications
Sample volume 25 µl of whole blood in three-part mode, 50 µl of whole blood in five-part mode. 50 µl
of whole blood in five-part prediluted mode of 1:6; 25 µl of prediluted blood in three-
part prediluted mode of 1:6.
Chambers 1 unified chamber for diluting whole blood and counting.
Reagent system VetScan HM5 Reagent pack.
Aperture diameter 80 µm.
Throughput 24–30 tests/hour in three-part mode, 16–20 tests/hour in five-part mode.
Sampling method Open tube system with automatic sample rotor, support for all HM2 and VetScan
HM5 test tubes.
Sample types Five-part mode: dog, cat, horse, alpaca, llama, cynomolgus macaque, rhesus
macaque, control. Dog2 and Cat2 are the dog and cat profiles, but may be adjusted
for puppy and kitten reference ranges, if desired.
Three-part mode: mouse, rat, rabbit, ferret, pig, cow, sheep, goat, guinea pig, primate
(primate is research only).
Clog prevention High-voltage pulse on aperture in each analysis cycle and chemical cleaning of the
aperture using Cleaner reagent.
Cleaning procedure High-voltage burn of the aperture, deep cleaning, chemical cleaning of the aperture.
Quality control Support for five-part differential control blood.
QC parameters include: mean, ± range, SD and CV for all measured and calculated
parameters, 16- and 64-day Levy-Jennings charts, separate QC database.
Calibration Three-measurement automatic calibration of WBC, HGB, RBC, PLT, MCV, RDWc,
MPV, and EOS absolute.
Multi-user feature Three-level multi-user operation with selective privilege levels, user identification with
ID and password. Contact Abaxis for more information.
User interface Easy-to-use, menu-driven touchscreen user interface with on-screen help.
Languages available English, French, German, Italian, Spanish, Czech, Polish, Portuguese, and Russian.
For other languages, contact Abaxis Technical Support — see page 1-3.
Data capacity 5000 results, including RBC, PLT, WBC three-part and five-part histograms on-board.
Data can be saved to USB drive or downloaded to computer.
Host computer interface Four USB A ports, one USB B port, and one Ethernet port. Support for ASCII-based
communication protocol only (V3.1).
Data back-up method Port for USB drive on side and back panels.
Software upgrade method USB drive.
External printer interface USB.
Built-in printer “Easy Paper Operation” built-in thermal printer.
Display 240x128-dot, high-contrast, backlit, graphics liquid crystal diode.
External keyboard Standard USB-compatible keyboard.
Power requirement 12V DC, 5 A, 60 W.
Power supply unit External, auto-ranging power unit for 100–120 or 200–240 VAC, 50–60Hz.
Operating temperature 59–86 °F (15–30 °C). Optimal temperature is 77 °F (25 °C).
Dimensions (W x D x H) 12.6 x 10.2 x 14.4 in (320 x 260 x 365 mm).
Net weight 27 lbs (12.3 kg).

Table of Contents

Other manuals for Abaxis VetScan HM5

Related product manuals